FREE SHIPPING
Ai Recirculating Chillers, 10 to 25°C
- Available in multiple capacities
- Across recirculating chillers have numerous safety mechanisms to keep you and your lab safe
- Designed for the perfect balance of economy and quality
$4,690
$4,455
and up
Image | SKU | Description | Filling Volume | Pumping Speed | Electrical | Quantity | Price | Add to Cart |
---|---|---|---|---|---|---|---|---|
|
WAC-1 | Ai Recirculating Chiller, 10 to 25°C | 17 L | 528 gal/hr | 110V, 60Hz, 1-PH, 1460W | 1 Each |
|
|
|
WAC-4 | Ai Recirculating Chiller, 10 to 25°C | 35 L | 33 L/min | 480V, 60Hz, 3-PH, 3.75kW | 1 Each |
|
Description
Across International's WAC series of recirculating water chillers are built for the perfect balance of quality and economy. These units are equipped with digital temperature controllers for precise adjustments. For mobility and ease-of-use, these chillers are manufactured with swivel casters.
The built-in stainless steel tank and water pump ensure durability and no-rust operation, even over long periods of heavy use. A water manifold is included so that your unit is immediately usable.
Note! Ambient temperature must be at or below 35°C (95°F) for chiller to operate effectively.
Safety features
The WAC series includes multiple safety mechanisms to protect your team, your lab, and your investment. Safety alarms and system halts are present for:
- Over current
- Compressor delay start
- Water pump
- Over load
- Built-in fuse
- Over-voltage and under-voltage
- Abnormal operations
- Phase protection
Specifications
SKU | WAC-1 | WAC-4 |
---|---|---|
Model | WAC-1 | WAC-4 |
Controller | Digital micro-chip controlled with LCD screen and fluid level display | Digital micro-chip controlled with LCD screen and fluid level display |
Electrical | 110V, 60Hz, 1-PH, 1460W | 480V, 60Hz, 3-PH, 3.75kW |
Cooling Capacity | 1935 kcal/h (Nominal cooling capacity is according to: Inlet chilled fluid temperature: 12°C, Outlet chilled fluid temperature: 7°C, Inlet cooling air temperature: 30°C, Outlet cooling air temperature: 35°C | 10366 kcal/h (Nominal cooling capacity is according to: Inlet chilled fluid temperature: 12°C, Outlet chilled fluid temperature: 7°C, Inlet cooling air temperature: 30°C, Outlet cooling air temperature: 35°C) |
Compressor Make | Panasonic, UL Listed | Danfoss (UL Listed) |
Compressor Power | 110VAC 60Hz 1-PH, LRA42, thermally protected | 460V 3-60Hz 8A 3750 watts, LR 25A, thermally protected |
Refrigerant | R410A (800g) | R407C (2,400g) |
Max Pressure | 10 bar. H(max); 20.5m, H(n): 16m | 22.6 bar, TS max: 50°C, TS min: -35°C |
Number of Compressors | One | One |
Water Pump Make | LX (CMFL2-20-A-W-G-BABE) | LX (CMF2-50T-A-W-G-BABE) |
Water Pump Power | 110VAC 60Hz 1-PH 370 watts 3.7A 3450 rpm IP55 | 480V 60Hz 3-PH 550 watts 2.5/Y1.4A 3450 rpm IP55 |
Pumping Speed | 528 gal/hr | 33 L/min |
Filling Volume | 17 L | 35 L |
Recommended Coolant | Distilled water | Distilled water |
Cooling Fan Make | Kunli (YWF A2S-300S-5DIIIA05) | Weiguang (YWF4D-350S-102/34-G) |
Cooling Fan Power | 110VAC 60Hz 1-PH 155 watts 1.35A 2700rpm 16uF/500V IP44 | 480V 60Hz 80/150 watts 0.27A 1640 rpm |
Temperature Accuracy | +/- 2 °C | +/- 2 °C |
Inlet Connection | KF40 flange | KF40 flange |
Outlet Connection | 8mm quick connect with ball valve | 8mm quick connect with ball valve |
Weight | 200 lbs | 375 lbs |
Dimensions | 29 x 17 x 35" | 41 x 22 x 54" |
Shipping Weight | 260 lbs | 485 lbs |
Shipping Dimensions | 36 x 21 x 42" | 50 x 25 x 61" |
Safety Features | Over current and compressor thermal protection | Over current, compressor delay start, water pump over load, built-in fuse, over-voltage and under- voltage, abnormal operations, phase protection |
Ambient Environmental Temp | Temperature range of chilled fluid: 5°C - 35°C. Temperature difference of chilled fluid between inlet & outlet: 3°C - 8°C. It is better to use the chiller while the ambient temperature at or below 35°C. | Temperature range of chilled fluid: 5°C - 35°C. Temperature difference of chilled fluid between inlet and outlet: 3°C - 8°C. It is better to use the chiller while the ambient temperature is at or below 35°C |
Compliance | CE compliant | CE compliant |